Cirrus HD-OCT for Retina, Glaucoma and Cornea by Carl Zeiss Meditec

Technical Specifications

OCT imaging Model 400 Model 4000
Methodology Spectral domain OCT
Spectral domain OCT
Optical source Superluminescent diode (SLD), 840 nm Superluminescent diode (SLD), 840 nm
Scan speed 27,000 A-scans per second 27,000 A-scans per second
A-scan depth 2.0 mm (in tissue),
1024 points
2.0 mm (in tissue),
1024 points
Axial resolution 5 μm (in tissue) 5 μm (in tissue)
Transverse resolution 15 μm (in tissue) 15 μm (in tissue)
Fundus Imaging Model 400 Model 4000
Methodology Live OCT Fundus™ technology
Line scanning ophthalmoscope
Live fundus image During alignment
During alignment and during
OCT scan
Optical source Superluminescent diode (SLD), 840 nm
Superluminescent diode (SLD), 750 nm
Field of view 36 degrees W x 21
degrees H
36 degrees W x 30 degrees H
Frame rate >0.7 Hz >20 Hz
Transverse resolution 45 μm (in tissue) 25 μm (in tissue)
Iris Imaging Model 400 Model 4000
Methodology CCD camera
CCD camera
Resolution 1280 x 1024 1280 x 1024
Live iris image During alignment During alignment
Electrical, Physical and Environmental Model 400 Model 4000
Weight 79 lbs (36 kg)
83 lbs (38 kg)
Dimensions of Instrument 26L x 17W x 21H (in)
65L x 44W x 53H (cm)
26L x 17W x 21H (in)
65L x 44W x 53H (cm)
Dimensions of Table 39L x 22W (in)
99L x 56W (cm)
39L x 22W (in)
99L x 56W (cm)
Fixation Internal, external Internal, external
Internal fixation focus adjustment -20D to +20D (diopters) -20D to +20D (diopters)
Input devices Keyboard, mouse Keyboard, mouse
Electrical rating (115V) Single Phase, 100-120V~ systems:50/60Hz, 5A Single Phase, 100-120V~ systems:50/60Hz, 5A
Fuse rating (115V) T 5A 250V T 5A 250V
Electrical rating (230V) Single Phase, 220-240V~ systems:50/60 Hz, 2.5A Single Phase, 220-240V~ systems:50/60 Hz, 2.5A
Fuse rating (230V) T 5A 250V T 5A 250V
Temperature (transport and storage) –40º to +70º C –40º to +70º C
Relative humidity (transport
and storage)
10% to 100%, including condensation 10% to 100%, including condensation
Atmospheric pressure (transport
and storage)
500 hPa to 1060 hPa 500 hPa to 1060 hPa
Temperature (operation) +10º to +35º C +10º to +35º C
Relative humidity (operation) 30% to 75%, excluding condensation 30% to 75%, excluding condensation
Atmospheric Pressure (operation) 700 hPa to 1060 hPa 700 hPa to 1060 hPa
Internal Computer Model 400 Model 4000
Processor Intel® Core™ 2 Duo Intel® Core™ 2 Quad
Memory 2GB 4GB
Hard drive / Internal storage ≥ 750 GB
> 80,000 scans
≥ 750 GB
> 80,000 scans
Display Integrated 15” color flat panel display Integrated 15” color flat panel display
CD/DVD DVD-Multi drive DVD-Multi drive
Software Features Model 400 Model 4000
Raster Scans Standard 5-Line Raster
Enhanced HD Raster
Macular Thickness and Macular Change Analysis X X
RNFL and ONH OU Analysis and Guided Progression Analysis X X
Normative Database – RNFL / ONH X X
Normative Database - Macula X X
Advanced Visualization™ and 3D display X X
Compatible with Cirrus Review Software (sold separately) X X
Anterior segment imaging (license may be sold separately) X X
Connectivity with ZEISS FORUM® Eye Care Data Management system X X